IOGP Well Control Incident Lesson Sharing

Unexpected failure of the Rotating Control Device bearing rubber during a Managed Pressure Drilling operation caused a loss of well control event


Note: Original alert wording from IOGP Member Company.

While the Bottom Hole Assembly (BHA) was being stripped to the casing shoe, a sudden and total failure of the MPD RCD bearing rubbers occurred thereby causing a rapid loss of the Surface Back Pressure (SBP). This resulted in a loss of the primary well control barrier (Fluid Hydrostatic Pressure + SBP). After approximately 20 minutes the drilling crew was able to reestablish well control by restoring the Equivalent Circulating Density (ECD) to an overbalance condition. The BHA was then pumped to the shoe and maintenance activities conducted to replace the RCD bearing and continue with normal operations.

The MPD system was installed with a statically underbalanced MW in hole.

The BHA was being pulled, stripping back to the shoe with surface back pressure against the MPD RCD prior to pumping 2.07sg tripping mud from the shoe to surface. A Surface Back Pressure of ~1000psi was applied.

At 4780m a sudden and total failure of the MPD RCD bearing rubbers occurred resulting in a loss of the SBP and as such a loss of the primary barrier (fluid+SBP) for ~20mins. The well was monitored on the active, while the failure was investigated, before re-establishing drilling ECD to restore overbalance.

Then the BHA was pumped back to the shoe to change the RCD bearing. No gain volume was recorded but two gas peaks of 8% and 5% were circulated.

What Went Wrong?:

  • Mechanical failure of the RCD bearing.
  • Delayed response due to previous experience of only partial failures (still able to hold SBP).
  • Confusion as to the best course of action due to - No significant reservoir drilled, no flow from the well on the active, tool joint passing the bearing at time of failure, sticking concern.

Corrective Actions and Recommendations:

  • Delayed response – clarify procedures for bearing failure, the expected reaction is to shut the well in on BOPs Enhance drills with the crews during fingerprinting to cover bearing failure.
  • Sudden and total failure of the RCD bearing can occur, the elements are particularly vulnerable under higher pressure stripping.
  • The contingency procedures within the MPD programme for a bearing leak scenario focused on a manageable leak (able to apply SBP) and did not include specific steps to take in case of total failure.
  • The operations team had not experienced a similar failure and therefore time was required to establish what had occurred. Note: No previous experience of sudden and total bearing failure.
  • In long and stable operations a shift of risk-ranking and priorities can occur subconsciously, e.g. being stuck vs. well control.
  • Full technical investigation of RCD element mechanical failure required. No cause identified yet.

Figure 1: RCD Bearing damaged - view 1

images

Figure 2: RCD Bearing damaged - view 2

images

Figure 3: Simplified Cause Tree Analysis

images


safety alert number: 317
IOGP Well Control Incident Lesson Sharing http://safetyzone.iogp.org/

Disclaimer

Whilst every effort has been made to ensure the accuracy of the information contained in this publication, neither the IOGP nor any of its members past present or future warrants its accuracy or will, regardless of its or their negligence, assume liability for any foreseeable or unforeseeable use made thereof, which liability is hereby excluded. Consequently, such use is at the recipient's own risk on the basis that any use by the recipient constitutes agreement to the terms of this disclaimer. The recipient is obliged to inform any subsequent recipient of such terms.
This document may provide guidance supplemental to the requirements of local legislation. Nothing herein, however, is intended to replace, amend, supersede or otherwise depart from such requirements. In the event of any conflict or contradiction between the provisions of this document and local legislation, applicable laws shall prevail.


IOGP